Our Products

TV400 Standard &TV400K Cooled Twin-Plate Maxi-Gel Electrophoresis Units

The TV400 and TV400K twin-plate 20.5 x 20cm (W x H) maxi-gel units offer all the benefits of the TV100 and TV200 units, but in a much larger gel size and with increased sample throughput. A cooling coil within the TV400K gel tank allows faster separations to be performed at high voltage without compromising resolution

Product Specification
Complete System Part No.
Twin-plate wide format mini-gel unit with GRM and gel tank, lid,
2 x (20.5 x 20cm; W x H) plain glass plates, 4 x 1mm spacers, 2 x spacer aligners, 2 x (20.5 x 20cm) notched glass plates, 1 x dummy plate and
2 x 1mm thick 24-sample combs
TV400
Twin-plate wide format mini-gel unit with GRM and cooled gel tank,
with built-in cooling coil and quick-fit tubing, plus lid, 2 x (20.5 x 20cm; W x H) plain glass plates, 4 x 1mm spacers, 2 x spacer aligners,
2 x (20.5 x 20cm) notched glass plates, 1 x dummy plate and 2 x 1mm thick 24-sample combs
TV400K
Twin-plate wide format mini-gel unit with GRM and gel tank, lid,
2 x (20.5 x 20cm; W x H) plain glass plates, 4 x 1mm spacers, 2 x spacer aligners 2 x (20.5 x 20cm) notched glass plates, 1 x dummy plate,
2 x 1mm thick 24-sample combs, casting base and 2 x silicone seals
TV400Y
Twin-plate wide format mini-gel unit with GRM and cooled gel tank, with built-in cooling coil and quick-fit tubing, plus lid, 2 x (20.5 x 20cm; W x H) plain glass plates, 4 x 1mm spacers,
2 x spacer aligners, 2 x (20.5 x 20cm) notched glass plates, 1 x dummy plate,
2 x 1mm thick 24-sample combs, casting base and 2 x silicone seals
TV400YK
Technical Specifications
Unit Dimensions (W x D x H) 28.5 x 15 x 27.5cm 38 x 15 x 27.5cm
Inner Tank Dimensions (W x D x H) 27 x 12.5 x 25cm 27 x 12.5 x 25cm
Plate Dimensions (W x H x T) 20.5 x 20 x 0.4cm 20.5 x 20 x 0.4cm
Standard Spacer Dimensions (W x H x T) 2 x 20 x 0.1cm 2 x 20 x 0.1cm
Active Gel Dimensions (W x H) 16.5 x 17.5cm 16.5 x 17.5cm
Maximum Sample Capacity 2 x 48 2 x 48
Recommended Buffer Volume Inner Buffer                700ml Inner Buffer                700ml
  Chamber Chamber
  Gel Tank                 4200ml Gel Tank                 4200ml
Recommended Running Conditions for Voltage                175 – 225V Voltage               200 – 300V
Denaturing/Native PAGE Gel (8.75 – 11.25V/cm) (10 – 15V/cm)
  Current              20 – 30mA Current              30 – 40mA
  Time                    2.5 – 3.5h Time                     2 – 2.5h
Snap-lock Connectors for Cooling Coil Inner Diameter                n/a Inner Diameter          10mm
  Outer Diameter               n/a Outer Diameter          12mm
Quick-fit Tubing Inner Diameter                n/a Inner Diameter          10mm
  Outer Diameter               n/a Outer Diameter          12mm
Casting Base Silicone Seal 1.7 x 22.5 x 0.8cm 1.7 x 22.5 x 0.8cm
Dimensions (W x L x H)    
Power Output Connectors (diameter) Shrouded, 4mm Shrouded, 4mm
Recommended Power Supply Consort EV243 Consort EV243

1 + 5 =

Other Products

SE600X Chroma™ Deluxe Dual Cooled Vertical Electrophoresis Unit

SE600X Chroma™ Deluxe Dual Cooled Vertical Electrophoresis Unit

Ergonomic design offers easier handling and sample introduction Cat. # Description SE600X-15-1.5 Deluxe Dual Cooled Vertical Electrophoresis Unit, Complete Includes: Lower Buffer Chamber Bubble Level Upper Buffer Chamber Laminated Gaskets for Caster–2 pcs Safety Lid...

Consort EV200 Range Power Supply Units

Consort EV200 Range Power Supply Units

The Consort EV200 power supplies cover the range of voltages and currents used with all Scie-Plas electrophoresis units, whether it is for routine low voltage horizontal and vertical gel electrophoresis or high voltage applications such as isoelectric focusing and...

Beta 3-Sided Shield

Beta 3-Sided Shield

The Beta 3-sided shield can be used with our range of Beta shields, providing further protection

CA-SYS Cellulose Acetate System

CA-SYS Cellulose Acetate System

The CA-SYS cellulose acetate system has been designed in close collaboration with the UK National Health Service (NHS) to run cellulose acetate membranes and wick-based electrophoresis applications, particularly in the two-dimensional screening of inherited metabolic...

The GD-4534 and GD-5040 Gel Dryers

The GD-4534 and GD-5040 Gel Dryers

A cast-aluminium base, overlaid with a porous stainless steel sheet, evenly distributes the heat generated by an 800-Watt heating element, which is set in increments of +/- 2°C to a maximum temperature of 90°C. Both the heating element and vacuum pump are...